Dear Sirs,
I have to mention one thing about my consult in the list: the dose rate that
we measured was at control booth. If we take the NCRP recommendations, the
dose in a wk must be less than 0,1 mGy. We calculate the dose in function of
the workload: the result was that the shielding is good. But usually, the
dose rate we measured is less than 100 uSv/h [kerma]. So, when we measured
the 500 uSv/h, we asked: is there a dose rate limit for practice purpose?
I hope this additional information will be useful to clear our trouble.
Thank you very much for who answered my first message.
God bless you,
Lic. Yuri Ravello
Evaluador-Inspector
Oficina Técnico de la Autoridad Nacional
Instituto Peruano de Energía Nuclear
Calle Justo Vigil 456 - Magdalena del Mar
Teléfono: +51.1.4631170 o +51.1.4631171
Fax: +51.1.4631166
-----Mensaje original-----
De: Yuri Ravello [mailto:yravello at ipen.gob.pe]
Enviado el: lunes, 27 de marzo de 2006 9:33
Para: 'radsafe at radlab.nl'
Asunto: Request information for shielding purpose
Dear Sirs,
We are testing a shielding design for a x-ray medical use equipment. We
measured the expose (kerma) rate: it is about 500 uSv/h. Because we cannot
measure integrated exposure, we are looking for the limit for expose rate.
So please, if anyone has experience about it, write me. Thanks in advance.