Doesn't anyone appreciate the energy response virtues and senstivity of
the tissue equivalent microrem/hr meter?
Mike Borisky
Army Research Lab
-----Original Message-----
From: radsafe-bounces at radlab.nl [mailto:radsafe-bounces at radlab.nl] On
Behalf Of Sam Iverstine
Sent: Friday, March 23, 2007 8:35 AM
To: radsafe at radlab.nl
Subject: RE: [ RadSafe ] GM vs ion chamber for dose measurement
Great topic. Great link to energy response curves.
One's employer could spend less on an ion chamber than the two days of
your time and large errors associated with applying curves and filters,
in my opinion. But it might be fun to overcome the challenge.
This brings me to my question for radsafe: What is the best hand held
ion chamber to purchase for mainly radiology uses? The best I can find
are the 450 and 451P's by Victoreen. Sometimes these meters give erratic
readings due to temperature or mechanical agitation (dropping).
They do have a good response rate and accurate digital readout over
all energies of interest. The plastic case is water-resistant and
filters fewer low energy photons (than a metallic encased meter). But
there must be something better in this millennia?
Thanks for any response.
Sam Iverstine, MS, CHP
Aad van der Kooij <A.vanderKooij at TUDelft.nl> wrote:
Well, by the time I received this message it was Friday ...
The response of a GM tube can be tuned by using filters around (parts
of)the tube. For example:
http://www.hps.org/hpspublications/articles/allard.html
And yes, you are right about questioning what is going on!
Dose rates are expressed per unit of time and(TGIF)dose in Sv so: Sv.h-1
-)
Enjoy your weekend!
Aad van der Kooij
RPO TU Delft (NL)
-----Original Message-----
From: radsafe-bounces at radlab.nl [mailto:radsafe-bounces at radlab.nl] On
Behalf Of Baumbaugh, Joel SPAWAR
Sent: Thursday, 22 March 2007 18:20
To: radsafe at radlab.nl
Subject: [ RadSafe ] GM vs ion chamber for dose measurement
Yes, I know this isn't Friday, but I have a philosophical question for
the group.
First a short introduction: My current employer appears to be
mandating/authorizing the use of Geiger Mueller (GM) detectors to
measure dose rates in Roentgens (the meters read "capital" R, not "r") .
As many of you know, a given instrument intended for performing
quantitative measurements will respond differently to various energies
of a given radiation, and a GM tube's response is quite energy
dependent. As the GM's are calibrated with Cs-137 the response to other
x-ray/gamma sources (of other energies/strength) can be quite different
than what one would measure with an ion chamber - i.e. the energy
dependence of the instrument must be known and accounted for in order to
make accurate quantitative measurements.
A short Discussion: I'm not sure that my employer realizes that by
utilizing a GM for measurement of (a wide energy-range of) gamma/x-rays
they are creating the potential for overexposure conditions (where in
some/many cases one should be utilizing an ion chamber). Exposure vs
dose, Bragg-Gray and kerma considerations put aside for now, I was
taught to always utilize an Ion Chamber if I REALLY wanted to know the
dose-rates (I use a pressurized ion-chamber myself for a wider range of
energy response). After all, last I knew, "exposure" was still defined
in terms of the amount of ionization charge created in AIR - and I
believe that the fundamental SI unit of exposure still corresponds to
that amount of x-or gamma radiation who's associated secondary electrons
create an ionization charge of on coulomb per kilogram of dry air at
STP. Anyone know of a GM that utilizes 78% nitrogen and 21% oxygen for a
counting gas?
So, in conclusion: Am I wrong to be questioning what's going on here? Is
the use of GM's for quantitative/qualitative measurement of gamma/x-ray
dose now an accepted protocol?
Thanks (ahead of time) for your thoughts.
Joel Baumbaugh
Std. Disclaimer... The thoughts, discussion and opinion(s) above are
mine and mine only, not that of my employer and, of course, I am in no
way asserting/suggesting that my employer is doing anything wrong. This
is only a topic for a philosophical Health Physics discussion and I
would value your opinion(s) on the matter.
_______________________________________________
You are currently subscribed to the RadSafe mailing list
Before posting a message to RadSafe be sure to have read and understood
the RadSafe rules. These can be found at:
http://radlab.nl/radsafe/radsaferules.html
For information on how to subscribe or unsubscribe and other settings
visit:
http://radlab.nl/radsafe/
_______________________________________________
You are currently subscribed to the RadSafe mailing list
Before posting a message to RadSafe be sure to have read and understood
the RadSafe rules. These can be found at:
http://radlab.nl/radsafe/radsaferules.html
For information on how to subscribe or unsubscribe and other settings
visit: http://radlab.nl/radsafe/
---------------------------------
Finding fabulous fares is fun.
Let Yahoo! FareChase search your favorite travel sites to find flight
and hotel bargains.
_______________________________________________
You are currently subscribed to the RadSafe mailing list
Before posting a message to RadSafe be sure to have read and understood
the RadSafe rules. These can be found at:
http://radlab.nl/radsafe/radsaferules.html
For information on how to subscribe or unsubscribe and other settings
visit: http://radlab.nl/radsafe/
Classification: UNCLASSIFIED
Caveats: NONE
More information about the RadSafe
mailing list