Does anyone have any insight so that I can get back to my researcher?

 

The contract specified a "6 ml polypropylene vials" for counting the 3H
samples. When I couldn't find those, and enquired, it turned out the
ones recommended were actually high density polyethylene.  There was a
specific recommendation from a particular vendor.  However, since we got
a much better deal on Vendor #2 HDPE 6 ml scintillation vials, we bought
those. 

 

The first samples we tried to count in those vials, with 1 ml of Scint
fluid (as recommended in the protocol), gave us < 10% of the expected
counts (925 instead of 12,000 DPM).  We were using the racks for the
mini-vials.  The technician then tried counting these in another
scintillation counter (the one used by Rad Safety), and they wouldn't
count at all:  this instrument gave error messages.  After determining
it wasn't a matter of bad pipetting or something, I had the technician
transfer that sample and its scintillation fluid to a HDPE 20 ml vial,
add 9 ml more scint. fluid, and he got the expected counts (just under
12,000).  I then had him take 5 ml of that mixture and put it back in a
clean 6 ml vial, and again, he got less than 10% of the counts expected
for the half of the volume counted (~500, instead of ~6000), so it
wasn't just the low volume of scintillation fluid suggested in the
protocol.  So it appears these vials don't count tritium well.   Do you
have any insight into this bizarre problem?  Vendor #2 has none, and
apparently has no data to actually support the use of these vials for
scintillation counting (at least for tritium, I didn't ask about other
isotopes).  They have offered to send me another type of vial, but that
doesn't entirely easy my mind.  

 

Any comments/experience with this sort of issue would be greatly
appreciated!!  Like, is it something that "everybody knows" (except me
and apparently the people that developed the protocol for the contract)
that one always has to count 3H in glass???   Have you ever heard of
HDPE vials contaminated with something that quenches the heck out of
tritium? 

 

 

Elaine T. Marshall, CHP