[ RadSafe ] Dose rate limitations
ROBBARISH at aol.com
ROBBARISH at aol.com
Thu Feb 10 23:04:40 CET 2005
Fellow RADSAFERS:
The current (2002 edition) "Medical and Dental Guidance Notes: A good
practice guide on all aspects of ionising radiation protection in the clinical
environment" used in Great Britain as the basis for regulatory compliance does
indeed contain instantaneous dose rate limits (IDR) for medical linear
accelerators of the type mentioned here. IDR is defined as being averaged over 1
minute.
According to the Guidance Notes, the IDR for a public area adjacent to the
primary barrier of a medical linear accelerator must be less than 7.5
microsievert per hour. (0.75 mrem/hr) This makes the typical primary barriers in
Great Britain a TVL thicker [about 18" (45 cm ) of concrete] than here in the
USA, where instantaneous dose limits are not used. (Don't get confused by the 2
mrem in-any-hour rule, which is not the same as 2 mrem per hour.) Other
values are established for controlled areas, but there are IDR limits everywhere.
As an example, a linear accelerator here with a total workload of 500 Gy per
week and a use factor of 1/4 toward a primary wall operating at 6 Gy per
minute might have a dose rate during a protection survey of 60 microsievert per
hour (6 mR/hr). The linac will be "on" pointing in that direction for
approximately 20 minutes per week (500 x 1/4)/600 = 21 minutes. So the total dose
will be 20 microsievert for the week; an acceptable 1 mSv for the year. In the
UK that dose rate would be 8 times too high, requiring the additional TVL of
shielding.
In Britain, the National Health Service foots the bill for the additional
protection, so it's always at taxpayers expense.
What's really interesting about this is the fact that new technologies might
come along that provide for higher instantaneous dose rates from the therapy
unit. Here, that would be irrelevant since, for example, doubling the dose
rate would, of course, half the treatment time so the integrated exposure
would remain constant.
There, the higher dose rate would require additional shielding. The document
states (on page 65) that treatment rooms should be designed considering
possible future developments including "increases in dose rates." So a shielding
designer for a British radiotherapy facility possibly has to have psychic
powers in order to predict, years in advance, the dose rates that might come
from future generations of medical accelerators.
Robert Barish, Ph.D. CHP
_robbarish at aol.com_ (mailto:robbarish at aol.com)
<Dear colleagues
<As we are installing now a new Clinac 2100C/D accelerator with 6 and 18
<MV X-rays , I would like to know if in your countries are there
<limitations on the max dose rate in the walls of the installation ,
<specifically for the direct beam walls. I am perfectly aware of the
<calculations for the max annual integral dose with the use factor,
<occupancy factor and so on. My question is related to the radiation
<protection survey of the competent authorities, and if there are
<guidelines concerning also the dose –rates measured for the max FS , for
<direct beam radiaton.
<Thank you in advance
<Sergio Faermann,Ph.D.
More information about the radsafe
mailing list