I agree with Franz. It is impossible to use a LSC to estimate the
tritium conc. without calibrating the system for the mixture used.
Quench can be a major factor. Anybody that has tried to and blood to a
cocktail knows that the efficiency goes to zero with very little added
blood. I would never accept LSC results without spiked sample for
calibration. I don't think quench measurements are good enough.