I'm looking for comments on the 2 methods for calculating
neutron mrem at the door to the maze in a linac room. (NCRP
79 pp 52-7)
I've done it using the Kersey method using a TVL
(tenth-value length) of 5 m in the maze, and got 370 & 385
mrem/wk (2 different locations for entrance to maze)
I then used the McCall's albedo method and got 98.7 mrem/wk.
The room geometry is close to Fig 34a (the "poor" one) on
p53, with a maze wall (room door to end) of just 7 feet.
Does this roughly 4 to 1 difference agree with others'
experiences??
What are other people doing ??
Thanks in advance.
--