Mark,
You want to look at 10 CFR 31.3 which allows up to 500 uCi per
device for the generally licensed static eliminators . They're
not included under 10 CFR 31.5 which requires the leak testing
you refer to.
Tim Popp
Pharmacia & Upjohn
timothy.l.popp@am.pnu.com
____________________________________________________
Date: Thu, 30 Apr 1998 08:54:12 -0600
From: mdater@nexstar.com
To: radsafe@romulus.ehs.uiuc.edu
Subject: Re: Static eliminator in copy machines
Message-ID: <872565F6.005066CE.00@pigeon.nexstar.com>
I have some questions regarding the static eliminators
(Staticmaster) for
microgram scales.
If they are 250 uCi ea. Why are they exempt? (exempt quantity is
100uCi or
less for 210Po) Don't they require leak testing?( sealed sources
containing
alpha emitting material >10 uCi do).
Mark Dater
NeXstar Pharmaceuticals, Inc.