[ RadSafe ] Re: attn. Landauer Customers
Sandy Perle
sandyfl at earthlink.net
Wed Aug 24 15:28:11 CDT 2005
On 24 Aug 2005 at 10:07, Neill Stanford wrote:
> the correct method is to subtract element specific values, not
> calculated dose.
Neill,
You are absolutely correct. Whenever possible, a dosimeter of the
same type, placed in the radiation free background area in the
facility that is monitored, subtracting the element response from the
individual's dosimeter elements is the correct process to be
followed. The only time a dose should be used is if there is no
actual control dosimeter. One must use something to account for
natural background as well as potential transit dose.
You're right on as usual!
Regards,
Sandy
----------------------------------------------------------------
Sandy Perle
Senior Vice President, Technical Operations
Global Dosimetry Solutions, Inc.
2652 McGaw Avenue
Irvine, CA 92614
Tel: (949) 296-2306 / (888) 437-1714 Extension 2306
Fax:(949) 296-1144
Global Dosimetry Website: http://www.dosimetry.com/
Personal Website: http://sandy-travels.com/
More information about the RadSafe
mailing list