[ RadSafe ] Scintillation Counting Question
Nielsen, Erik
nielseec at nv.doe.gov
Wed Aug 27 10:18:57 CDT 2008
To expand on Cary's comment.
I would recommend vigorous shaking of your sample/cocktail mixture and
degassing it by setting it in an ultrasonic cleaning bath for a few
minutes. The ultrasound will degas the sample as well as contribute to
the homogenization of the sample/cocktail mixture.
Erik C. Nielsen
Senior Scientist
Remote Sensing Laboratory
P.O. Box 98521, M/S RSL-47
Las Vegas, NV 89193-8521
http://www.nv.doe.gov/nationalsecurity/homelandsecurity/frmac/default.ht
m
Voice 702-295-8954
Fax 702-794-1007
Pager 702-794-7222
Cell 702-630-2323
Text to Pager 702794722 at nv dot doe dot gov (plain text please)
Text to Cell: 7026302323 at mmode dot com (plain text please)
This email message is intended only for the use of the named recipient.
Information contained in this email message and its attachments may be
privileged, confidential and protected from disclosure. If you are not
the intended recipient, please do not read, copy, use or disclose this
communication to others. Also please notify the sender by replying to
this message and then delete it from your system.
========================================================================
===============================
Message: 8
Date: Tue, 26 Aug 2008 17:41:38 -0700
From: "Cary Renquist" <cary.renquist at ezag.com>
Subject: RE: [ RadSafe ]
To: "Marshall, Elaine" <emarshall at lrri.org>, <radsafe at radlab.nl>
Just a thought...
Perhaps it is a physical phenomenon.
Since the tritium compound is moderately soluble, perhaps it initially
forms
small (microscopic) globules in the aqueous solution that eventually
"dissolve"
in the cocktail -- the globules might not be large enough to see, but
large enough
to keep the tritium beta from interacting with the cocktail.
Cary
-----Original Message-----
From: radsafe-bounces at radlab.nl [mailto:radsafe-bounces at radlab.nl] On
Behalf Of Marshall, Elaine
Sent: Monday, 25 August, 2008 14:47
To: radsafe at radlab.nl
Subject: [ RadSafe ] Scintillation Counting Question
I have a researcher who posed the following question to me. I have not
experienced this yet and was wondering if anyone else might have an
idea.
"For this study counting tritium (as estradiol), we are using Ultima
Gold cocktail, 10-12 ml for 40 microliter samples. We are seeing
something that seems odd to me: after holding for at least half an hour
in the dark, we get counts that seem low, and if re-counted a day or so
later, the counts go UP, by about 50%, and reach the expected levels. I
know of a number of reasons for counts to go DOWN, but don't understand
how they can go up. We do make a point of mixing well, without shaking
so much as to cause bubbles. Any clue???"
The researcher wanted me to add that these are aqueous samples, although
the actual tritiated compound is only moderately aqueous-soluble.
Thanks!!
Elaine T. Marshall, CHP
Radiation Safety Officer
Lovelace Respiratory Research Institute
2425 Ridgecrest Drive SE
Albuquerque, NM 87108
(505) 348-9578
emarshall at lrri.org
------------------------------
More information about the RadSafe
mailing list