[ RadSafe ] CT incident - how could it happen? Too many slices, high skin dose

parthasarathy k s ksparth at yahoo.co.uk
Wed Nov 12 19:46:25 CST 2008


Don't you think that 151 slices are too many! Skin dose in CT under certain conditions could be a few tens of mGy.

K.S.Parthasarathy




________________________________
From: "garyi at trinityphysics.com" <garyi at trinityphysics.com>
To: radsafe at radlab.nl; Mauro Campoleoni <trentino at iol.it>
Sent: Wednesday, 12 November, 2008 7:39:46
Subject: Re: [ RadSafe ] CT incident - how could it happen?

It seems extremely unlikely.  The article, including Brenner's remarks, are pretty "nuanced"
and don't actually say anything definite about the exam or the radiation dose delivered.  It is
possible that the "151 scans" are actually 151 *slices* of 1 mm each, acquired in one or two
passes with a multislice scanner.  The article may simply be designed to put public pressure
on the hospital to $ettle quickly.

If dose implied by the "red marks all the way around his head, like a severe sunburn"
statement were true, the child would be dead or very sick.  At least, if the erythema is
radiation induced there should be some epilation also, but there is no mention of that in the
article.

I wonder if it is possible to do EPR dosimetry on human hair?

-Gary Isenhower


On 12 Nov 2008 at 15:01, Mauro Campoleoni wrote:

From:               "Mauro Campoleoni" <trentino at iol.it>
To:                 <radsafe at radlab.nl>
Date sent:          Wed, 12 Nov 2008 15:01:54 +0100
Subject:            [ RadSafe ] CT incident - how could it happen?

[ Double-click this line for list subscription options ]

Hallo friends,

does anybody know about this?
http://www.diagnosticimaging.com/news/article/113619/1348813

Any idea of how it could happen (if it really happened)?

Thanks!

Mauro Campoleoni
----------------------------------------------------------------------
------ Unità Operativa di Fisica Sanitaria Fondazione "Ospedale
Maggiore Policlinico Mangiagalli e Regina Elena" Via Pace, 9 20122 -
Milano - ITALY tel. 02-5503.3007 fax 02-5503.5100
----------------------------------------------------------------------
------


_______________________________________________
You are currently subscribed to the RadSafe mailing list

Before posting a message to RadSafe be sure to have read and
understood the RadSafe rules. These can be found at:
http://radlab.nl/radsafe/radsaferules.html

For information on how to subscribe or unsubscribe and other settings
visit: http://radlab.nl/radsafe/
_______________________________________________
You are currently subscribed to the RadSafe mailing list

Before posting a message to RadSafe be sure to have read and understood the RadSafe rules. These can be found at: http://radlab.nl/radsafe/radsaferules.html

For information on how to subscribe or unsubscribe and other settings visit: http://radlab.nl/radsafe/



      


More information about the RadSafe mailing list