[ RadSafe ] Best technique to detect DU in huge area!!!
Dixon, John E. (CDC/ONDIEH/NCEH)
gyf7 at cdc.gov
Fri Nov 19 14:38:10 CST 2010
A Fidler is a special, thin NaI detector made for detection of Plutonium or Americium. A 2x2 inch NaI detector, not a Fidler, should do since you will be detecting the U-238 daughter Pa-234 (about 1 Mev).
John Dixon
----- Original Message -----
From: radsafe-bounces at health.phys.iit.edu <radsafe-bounces at health.phys.iit.edu>
To: radsafe at health.phys.iit.edu <radsafe at health.phys.iit.edu>
Sent: Fri Nov 19 12:22:23 2010
Subject: [ RadSafe ] Best technique to detect DU in huge area!!!
Dear All
Your advice is greatly appreciated:
There is a potential to perform Radiation survey in huge area 10 km x 30 km
or more for possible contamination with DU as a result of Gulf war.
The area is desert, sandy and not uniform.
What is the best technique to perform that?
Is movable FIDLER (NaI 2inx2in) with GPS and data timing is the best?
Any advice about the cost (individually and array as example 6 with trailer)
Suitable manufacturer and contact. for complete system..
Best regards
Dr. Falah Abu-Jarad
Dr.fabujarad at gmail.com