Thompson, Dewey L DThompson3 at ameren.com
Sun Jun 1 09:28:07 CDT 2014


The most common method I've seen is to use a depleted Uranium slab. Due to the energies involved with all of the betas emitted and electronic equilibrium, the surface dose rate is a constant, usually taken at 230 mRem (2.3 mSv)/hr. 

Some facilities use a specific isotope, for example P-32, (and I don't remember the dose rates on that).

The beta factor is simply the known dose rate divided by the open - closed value. 

For a uranium slab with commonly available hand helds (large window), it wi typically be between 3 and 5. Often for field use they will use a standard multiplier of 4. 

Dewey
Sent from my iPhone

> On Jun 1, 2014, at 7:05 AM, "Brent Rogers" <brent.rogers at optusnet.com.au> wrote:
> 
> On a similar vein, does anyone know for certain how the beta-factor on ion chambers is determined?  I'm speaking of obtaining beta dose rates using (open window - closed window) * beta factor.
> 
> Regards
> Brent Rogers
> Sydney Australia
> 
> Sent from my iPad
> 
>> On 1 Jun 2014, at 3:55 pm, ROY HERREN <royherren2005 at yahoo.com> wrote:
>> 
>> 
>> The calibration procedures will vary depending on which specific instrument your calibrating.  Unless you have the in-house expertise, I highly suggest that you leave the instrument calibration to the instrument vendor.  In any case, your instrument vendor should have provided you with an instrument manual when you purchased each of the instruments.  Look through the instrument manual to see if it lists any information about instrument calibration.  Do you own an instrument calibration source?  If so, also look up what the literature from calibration source manufacturer/vendor provides regarding instrument calibration. 
>> 
>> 
>> Roy Herren
>> 
>> 
>> On , ROY HERREN <royherren2005 at yahoo.com> wrote:
>> 
>> 
>> 
>> 
>> Are you specifically referring to the Rad MonitorTM instruments manufactured by Lab Logic, see Rad-Monitors
>> 
>> 
>>  Rad-Monitors  
>> View on www.lablogic.com Preview by Yahoo  
>> 
>> 
>> Roy Herren
>> 
>> 
>> On Friday, May 30, 2014 1:18 PM, "JOHN.RICH at sargentlundy.com" <JOHN.RICH at sargentlundy.com> wrote:
> _______________________________________________
> You are currently subscribed to the RadSafe mailing list
> 
> Before posting a message to RadSafe be sure to have read and understood the RadSafe rules. These can be found at: http://health.phys.iit.edu/radsaferules.html
> 
> For information on how to subscribe or unsubscribe and other settings visit: http://health.phys.iit.edu
> 

The information contained in this message may be privileged and/or confidential and protected from disclosure. If the reader of this message is not the intended recipient, or an employee or agent responsible for delivering this message to the intended recipient, you are hereby notified that any dissemination, distribution or copying of this communication is strictly prohibited. Note that any views or opinions presented in this message are solely those of the author and do not necessarily represent those of Ameren. All e-mails are subject to monitoring and archival. Finally, the recipient should check this message and any attachments for the presence of viruses. Ameren accepts no liability for any damage caused by any virus transmitted by this e-mail. If you have received this in error, please notify the sender immediately by replying to the message and deleting the material from any computer. Ameren Corporation




More information about the RadSafe mailing list