H-3 Bremsstrahlung - Re: G. Kramer
Measured H-3 bremsstrahlung ?
Shure we have Gary !
This is one of Ric Surette's favorite subjects (Ref.
Surette, R.A. A novel tritium process monitor that uses an X-ray detector,
Nuclear Instr. & Methods A 337 (1994) 588-593).
We can measure tritium contaminated (highly contaminated) stainless steel with
a thin window pancake detector. Infact, we show this in our annual tritium safe
handling course. However, at levels of H-3 contamination that
most RSOs are likely to run into an open window GF proportional counter, or
swipe-LSC is required to detect the tritium directly. The bremsstrahlung yield
from tritium interacting with an absorber is of the order 10E4 bremsstrahlung
X-rays per tritium betas decay.
Mike Wood
Radiation Biology and Health Physics Branch
Chalk River Laboratories