[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]
Re: BF3 Neutron Detectors
>>If you're going to use BF3 (vs He3) the stainless steel tubes from LND (I
>>think) seem to be among the best we've tried. We have not had great
>>results
>>from the N. Wood tubes. Stainless tubes are of course more expensive.
Actually from LND the stainless is only SLIGHTLY more expensive.
We also recently ordered a He3 tube from LND and had I have had a chance to
check it out. The results are VERY encouraging!! The He tube is 3.5 times
more sensitive and just as good at gamma rejection - the BEST part is that
it operates 225 volts lower (1225 instead od 1450!!). I am planning on
switching entirely to He3 as soon as I can get the $ to do so!
I was originally concerned because of conflicting information on gamma
rejection - but it is simply no problem!
I'd also like to order one with internal an internal alpha source of around
500 dpm. I'd use this for constancy checks/evaluation of the electronics.