Judd Sills wrote:
snip.....
> Electroplated sources are good for this type of situation. FYI, the Ludlum
> reported efficiency specifications (2 Pi) for this beta-gamma pancake probe are:
> C-14 10%
> Sr-90/Y-90 45%
> Tc-99 38%
> P-32 65%
> Pu-239 30%
>
> The 4 Pi geometries are one half of the above values.
>
>
Dear Radsafers,
Judd makes some very good points. Thanks.
Looking at the above efficiencies, there is an apparent inconsistency.
Sr-Y-90 should have a substantially higher counting efficiency than P-32
since Sr-Y-90 has 2 betas per decay, one of which (2.27 Mev) is quite a
bit higher energy than that of P-32, and the other (0.546 MeV) ain't too
shabby either.
Are Ludlum's specs just wrong here, or am I missing something?
Wes
--