[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]

Re[2]: 137Cs check sources -Reply



I thought, and have taught several thousand students, that the problem was
because of the long resolving time of the GM tube itself.  The tube takes
several hundred microseconds to recover from the electron cascade and is
unable to detect a second event during that time.  If this is not the case,
I'll change my tune.

Thanks in advance.
************************************************************************
The RADSAFE Frequently Asked Questions list, archives and subscription
information can be accessed at http://www.ehs.uiuc.edu/~rad/radsafe.html