Re: Discrpancies among different dosimeters



At 05:58 PM 5/19/99 -0500, you wrote:
>I am trying to characterize radiation field from 100 ns x-ray pulse.

Sandy points out that the differences are probably related to the energy
spectrum. However, ...

>I did not pay attention to whether the dosimeters were facing the beam 

It seems to me that this is a very important variable. The actual TLD
material is thin, so the electron equilibrium is determined by the
surrounding materials. Sandy, please correct me: I assume the metal clip of
a TLD holder would have a large effect if it were situated between the
x-ray source and the TLD chip.

mike
************************************************************************