[Date Prev][Date Next][Thread Prev][Thread Next][Date Index][Thread Index]
Re: N13.11-1993
> b) An irradiation level is determined at random using Appendix A, Equation A1
> for each dosimeter (in this case ensuring that no more than 1 TLD is irradiated
> to less than 50 mrem).
The above is the appropriate method. However, be aware that the lower
dose threshold is 30 mrem for Cat. IV, and there amy be more than 1
under 50 mrem within the 3 months tested (15 total dosimeters per
Category)
Sandy Perle
E-Mail: sandyfl@earthlink.net
Personal Website: http://www.geocities.com/capecanaveral/1205
"The object of opening the mind, as of opening
the mouth, is to close it again on something solid"
- G. K. Chesterton -
************************************************************************
The RADSAFE Frequently Asked Questions list, archives and subscription
information can be accessed at http://www.ehs.uiuc.edu/~rad/radsafe.html
- References:
- N13.11-1993
- From: "JOHN L. MCCARTHY" <MCCARJL@GWSMTP.NU.COM>