PCM-2 turning off detectors by itself yet showing "green" light (operational)
Greetings!
We had an interesting problem happen recently and I thought I would check
with the Radsafe community to see if anyone's had a similar problem. A
PCM-2 apparently disabled all 34 alpha and 34 beta channels by itself, yet
it still appeared to the workers that it was operational, i.e., had the
green light. It would go through the motions of performing a personnel
survey and indicate that the person had nothing over the limits on them,
but all detectors were inactive. The Instrumentation RCT found the PCM in
this disabled state when he went to perform a source check on it. He then
went to pull the PCM log data at my request so we could determine the
approximate time frame this happened. After pulling the data file
(involves exiting the PCM program to DOS), the RCT restarted the PCM
program and the unit went back to the state it was in prior to all
channels being turned off. We've investigated this and have ruled out
operator error and sabotage. The high-level password required to access
the menus to turn off the detectors had just been changed and only three
Instrumentation RCTs had it.
BNFL had Thermo Eberline modify the PCM program last year so that a tech
couldn't accidentally turn off all the detector channels with a single
keystroke. This accidentally happened last year when one of the field
RCTs, trying to help out, accidentally did this. The menu options to turn
off a single detector channel and to turn off ALL detector channels are
right next to each other in the menu and the screens for each of them look
exactly alike, except one says "all detector channels." The modification
Thermo made for BNFL requires that each detector channel has to be
individually turned off, so someone would have to send a lot of time to
turn off a total of 68 channels. The last incident was issued as a
"Lessons Learned" through the DOE system, which warned that the PCMs will
still show a green light (i.e., appears to be operational) even if all
detectors are off.
Anyone seen anything similar to this?