Re: Harshaw TLD info



The most efficient way to separate a TLD 100 from a TLD 700 is to 
irradiate them to low energy beta (i.e., Tl-204). The TLD-100 will 
under-respond significantly as compared to the TLD 700. A relative 
response correction for Tl-204 is approximately 1.5 (TLD 700) 
compared to approximately 5.5 for the TLD 100.


------------------