There were several good comments, and I hesitate to offer the following simple thoughts. I
noticed that there was a lot of variation in the dose measurements. If there were some
steady buildup of deposits, or deterioration of the anode, I would expect a more regular
decline in the measured dose. This makes me think that the generator, or perhaps the
power supply, may be the problem. A rigorous test of reproducibility will either eliminate or
confirm that. Several tests over the course of a couple of days may be called for.
-Gary Isenhower
************************************************************************